说:1.现在uv-100在我们办公室里使用,来测量空气,没有整合到系统中;2.普通空气,没有加压;3.如下
-186.76,1.9,971.3,813,0.0000,30/03/17,00:19:49
-13.17,1.9,970.2,812,0.0000,30/03/17,00:19:59
129.75,1.9,969.7,793,0.0000,30/03/17,00:20:09
44.68,1.9,969.9,820,0.0000,30/03/17,00:20:19
195.61,1.9,969.8,807,0.0000,30/03/17,00:20:29
365.36,1.9,969.5,790,0.0000,30/03/17,00:20:39
227.89,1.9,970.4,767,0.0000,30/03/17,00:20:49
198.98,1.9,969.9,871,0.0000,30/03/17,00:20:59
255.96,1.9,969.6,759,0.0000,30/03/17,00:21:09
132.38,1.9,970.3,809,0.0000,30/03/17,00:21:19
1.78,1.9,969.7,766,0.0000,30/03/17,00:21:29
-5.17,1.9,970.6,881,0.0000,30/03/17,00:21:39
17-03-31 14:05
附表中的第二列显示1.9K的恒定温度,当然不准确。 所以温度传感器在过去一年里已经被拔掉或损坏了。
我们建议您卸下顶部面板,并通过附带的PDF(uv-100下载中有)确认温度连接器已正确插入电路板。 如果它到位,那么它可能会被损坏,在这种情况下,我们可以给你一个新的。