The centrifuge speed can be set to a minimum of 300 rpm. You can adjust the speed in 100 rpm
increments. Proceed as follows:
1. Press one of the set keys (˄ for increase or ˅ for decrease) in the field of speed of the control
panel.
2. If you hold the key pressed, the display changes continuously at first slowly, and after a few
seconds at an accelerated pace to the higher or lower values.
3. Release the key as soon as you have reached the desired value and fine tune if necessary by
repeatedly briefly pressing on of the keys. The first digit behind the decimal point flashes for a
few seconds and then switches permanently into the current value mode. The new pre‐selected
speed is now stored.
Entering the RCF value
You can adjust the RCF pre‐selected value in steps of 100 g. The RCF value can be set to a minimum of
100 x g. You can adjust the RCF values the same as you adjust the rpm values. The displayed RCF value is
always corresponding to the maximum of centrifuge radius of the Micro liter rotor 24 x 2mL. For other
rotors please calculate the RCF value.
Selecting the run time
You can select a run time between 1 and 99 minutes or continuous operation [hd].
Run time selection
1. Press one of the “set” keys (˄ for increase or ˅ for decrease) in the field of